HPLC baseline noise

118 views 5 m ago

When using Acetonitrile and water composition in the ratio 50:50 premixed and sonicated for 10mins and ran one injection of blank,it's look like noisy baseline.so using nylon filter and used still te noise occurs.
I used this mobile phase in RS test with the run time of 70mins .

Give me a suggestion to reduce the noise by using this mobile phase.

1 Answers
Your Answer
Log in to answer this question.
1
1. Check column if there is any bubble, keep the open from outlet and rotate it and let the mobile phase pass for 10-15 mins.
2. Give back pressure from the outlet of Detector.
3. Equilibrate system for atleast 1 hr.